As far as your request is concerned, generally speaking the entire range of CDR instruments are
supplied already calibrated, ready to use and they do not need periodical calibration.
Anyway, two options are available to verify the calibration and the general functionality of the
system:
1. Hardware photometric check.
This procedure allows to verify the status of the instrument: photometric calibration, correct
functionality of parameters and reading cells, by following the procedure here below.
- Connect the instrument with an external PC, by using the USB cable provided as instrument
accessory.
- Now the CDR Drive folder will be created on the PC.
- From main menu, select SETTINGS.
- Select EXPORT STATUS and find the diagnostic files created in the CDR Drive folder.
- Send all the files to CDR that will check the status of the instrument and will release the updated
Certificate of Calibration.
2. Analytical check.
Please
write tosupport@cdrfoodlab.com, describing the problem in details,
with the help of pictures and brief videos, if needed.
>The instrument periodically asks for a new upgrading, just to remember it to the user and avoid to let use older versions. It happens also, as in this case, the instrument is already upgraded to the latest software version.
This happens when the
main switch is pressed for a longer time, and the instrument accesses to a
service menu called DEVICE UPGRADE MODE.
Please switch off the
instrument and repeat the procedure.
You have to press the main switch just for a while, one single touch, and the instrument will be switched on normally.
You can export results by following the procedure here below.
Moreover, you can download the software interface CDR Data Manager from the dashboard in this site https://resources.cdrfoodlab.com/, for a smarter way to manage your results.
From the instrument main
menu, please select SETTINGS and then ANALYSIS METHODS.
The same instruction,
including indication about sample treatment and sample dilution when required, can
be consulted after the method has been selected,by selecting the HELP key on
the screen.
Once selected the CDR instrument model on the home page, it is possible to submit the request Certificates of Analysis, and MSDS of CDR reagents.
Please check the code for which the request has been submitted.
CDR does not provide Certificates of Analysis/MSDSs for single components of a kit.
CDR provides Certificates of Analysis/MSDSs for complete kits only, please submit the request for 7 digits codes only.
CDR declares that the products, for which a
refrigerated storage is required, can be kept to superior temperature for some
days in order to guarantee the transport procedures and delivery to the final
customer.
We recommend the following indications:
- Products, for which storage at -20°C (-4°F) is
required
o
can be kept at temperatures from 2°C to 8°C (35.6°F – 46.4°F) for a maximum
of 30 days
o
can be kept at temperatures from 8°C to 25°C (46.4°F – 77°F) for a maximum
of 5 days.
- Products, for which storage at 2°C - 8°C (35.6°F –
46.4°F) is required,
o
can be kept for up to 10 days at temperatures not exceeding 25°C (77°F).
For longer time and temperature more than 25°C (77°F) it is not possible to
guarantee the expiring date even if the product has been well stored after
delivery.
Expired reagents cannot be used, CDR cannot guarantee the performance of reagents once they are expired.